Open Access. Powered by Scholars. Published by Universities.®

Medicine and Health Sciences Commons

Open Access. Powered by Scholars. Published by Universities.®

Western University

Oncology

Radiotherapy Planning

Publication Year

Articles 1 - 6 of 6

Full-Text Articles in Medicine and Health Sciences

Schedule For Ct Image Guidance In Treating Prostate Cancer With Helical Tomotherapy., G Beldjoudi, S Yartsev, G Bauman, Jerry J. Battista Dr., Jacob Van Dyk Mar 2010

Schedule For Ct Image Guidance In Treating Prostate Cancer With Helical Tomotherapy., G Beldjoudi, S Yartsev, G Bauman, Jerry J. Battista Dr., Jacob Van Dyk

Oncology Publications

The aim of this study was to determine the effect of reducing the number of image guidance sessions and patient-specific target margins on the dose distribution in the treatment of prostate cancer with helical tomotherapy. 20 patients with prostate cancer who were treated with helical tomotherapy using daily megavoltage CT (MVCT) imaging before treatment served as the study population. The average geometric shifts applied for set-up corrections, as a result of co-registration of MVCT and planning kilovoltage CT studies over an increasing number of image guidance sessions, were determined. Simulation of the consequences of various imaging scenarios on the dose …


The Effects Of Field-Of-View And Patient Size On Ct Numbers From Cone-Beam Computed Tomography., Katrina Y T Seet, Arvand Barghi, Slav Yartsev, Jacob Van Dyk Oct 2009

The Effects Of Field-Of-View And Patient Size On Ct Numbers From Cone-Beam Computed Tomography., Katrina Y T Seet, Arvand Barghi, Slav Yartsev, Jacob Van Dyk

Oncology Publications

Cone-beam computed tomography (CBCT) is used for patient alignment before treatment and is ideal for use in adaptive radiotherapy to account for tumor shrinkage, organ deformation and weight loss. However, CBCT images are prone to artifacts such as streaking and cupping effects, reducing image quality and CT number accuracy. Our goal was to determine the optimum combination of cone-beam imaging options to increase the accuracy of image CT numbers. Several phantoms with and without inserts of known relative electron densities were imaged using the Varian on-board imaging system. It was found that CT numbers are most influenced by the selection …


Adaptive Radiation Therapy For Localized Mesothelioma With Mediastinal Metastasis Using Helical Tomotherapy., James Renaud, Slav Yartsev, A Rashid Dar, Jacob Van Dyk Jan 2009

Adaptive Radiation Therapy For Localized Mesothelioma With Mediastinal Metastasis Using Helical Tomotherapy., James Renaud, Slav Yartsev, A Rashid Dar, Jacob Van Dyk

Oncology Publications

The purpose of this study was to compare 2 adaptive radiotherapy strategies with helical tomotherapy. A patient having mesothelioma with mediastinal nodes was treated using helical tomotherapy with pretreatment megavoltage CT (MVCT) imaging. Gross tumor volumes (GTVs) were outlined on every MVCT study. Two alternatives for adapting the treatment were investigated: (1) keeping the prescribed dose to the targets while reducing the dose to the OARs and (2) escalating the target dose while maintaining the original level of healthy tissue sparing. Intensity modulated radiotherapy (step-and-shoot IMRT) and 3D conformal radiotherapy (3DCRT) plans for the patient were generated and compared. The …


Consistency Check Of Planned Adaptive Option On Helical Tomotherapy., M Schirm, S Yartsev, G Bauman, Jerry J. Battista Dr., Jacob Van Dyk Dec 2008

Consistency Check Of Planned Adaptive Option On Helical Tomotherapy., M Schirm, S Yartsev, G Bauman, Jerry J. Battista Dr., Jacob Van Dyk

Oncology Publications

This study aims to evaluate a new Planned Adaptive software (TomoTherapy Inc., Madison, WI) of the helical tomotherapy system by retrospective verification and adaptive re-planning of radiation treatment. Four patients with different disease sites (brain, nasal cavity, lungs, prostate) were planned in duplicate using the diagnostic planning kVCT data set and MVCT studies of the first treatment fraction with the same optimization parameters for both plan types. The dosimetric characteristics of minimum, maximum, and mean dose to the targets as well as to organs at risk were compared. Both sets of plans were used for calculation of dose distributions in …


Quality Assurance Of Radiation Therapy Planning Systems: Current Status And Remaining Challenges., Jacob Van Dyk May 2008

Quality Assurance Of Radiation Therapy Planning Systems: Current Status And Remaining Challenges., Jacob Van Dyk

Oncology Publications

Computerized radiation therapy planning systems (RTPSs) are pivotal for treatment planning. The acceptance, commissioning, and quality control of RTPSs are uniquely complex and are described in the American Association of Physicists in Medicine Task Group Report 53 (1998) and International Atomic Energy Agency Technical Report Series No. 430 (2004). The International Atomic Energy Agency also developed a document and data package for use by vendors and purchasers to aid with acceptance testing of RTPSs. This document is based on International Electrotechnical Commission standard 62083 (2000) and describes both "type" tests to be performed in the factory and "site" tests to …


Adaptive Radiotherapy Planning On Decreasing Gross Tumor Volumes As Seen On Megavoltage Computed Tomography Images., Curtis Woodford, Slav Yartsev, R. Dar, Glenn Bauman, Jacob Van Dyk Nov 2007

Adaptive Radiotherapy Planning On Decreasing Gross Tumor Volumes As Seen On Megavoltage Computed Tomography Images., Curtis Woodford, Slav Yartsev, R. Dar, Glenn Bauman, Jacob Van Dyk

Oncology Publications

PURPOSE: To evaluate gross tumor volume (GTV) changes for patients with non-small-cell lung cancer by using daily megavoltage (MV) computed tomography (CT) studies acquired before each treatment fraction on helical tomotherapy and to relate the potential benefit of adaptive image-guided radiotherapy to changes in GTV.

METHODS AND MATERIALS: Seventeen patients were prescribed 30 fractions of radiotherapy on helical tomotherapy for non-small-cell lung cancer at London Regional Cancer Program from Dec 2005 to March 2007. The GTV was contoured on the daily MVCT studies of each patient. Adapted plans were created using merged MVCT-kilovoltage CT image sets to investigate the advantages …