Full-Text Articles in Medicine and Health Sciences

Characterization Of Treatment Planning System Photon Beam Modeling Errors In Iroc Houston Phantom Irradiations, Mallory Glenn Aug 2020

Characterization Of Treatment Planning System Photon Beam Modeling Errors In Iroc Houston Phantom Irradiations, Mallory Glenn

Dissertations & Theses (Open Access)

In radiation therapy, proper commissioning of the treatment planning system’s (TPS) dose calculation algorithm is critical because any errors in this process impact all treatment plans prepared in the system. Previously, TPS errors have been identified as a major cause for poor phantom irradiation performance, which may also mean that patients are treated suboptimally. The purpose of this work was to investigate the TPS beam modeling developed by the radiotherapy community to understand where inconsistencies may arise, which variables are most susceptible to variations, and in what way changing these variables can alter dose calculations.

Using the Imaging and Radiation …


Commissioning Of Micro-Cube Thermoluminescent Dosimeters For Small Field Dosimetry Quality Assurance In Radiotherapy, Brandon Luckett Aug 2019

Commissioning Of Micro-Cube Thermoluminescent Dosimeters For Small Field Dosimetry Quality Assurance In Radiotherapy, Brandon Luckett

Dissertations & Theses (Open Access)

Small field dosimetry presents complications and uncertainties that could be circumvented by using detectors which are smaller than the radiation field. This study evaluates the reproducibility and accuracy of TLD micro-cubes for use in stereotactic radiosurgery (SRS) remote auditing quality assurance (QA) for treatment centers participating in clinical trials. This study tested the hypothesis that TLD micro-cubes could be commissioned to evaluate small field dosimetry, and provide reproducibility within 3%, as well as assure agreement between measured dose and calculated doses to within 5%.

The aims of this thesis were to characterize and commission TLD micro-cubes as well as to …


Quality Assurance Of Advanced Treatment Modalities Using Presage® Dosimeters, Ryan G. Lafratta Aug 2015

Quality Assurance Of Advanced Treatment Modalities Using Presage® Dosimeters, Ryan G. Lafratta

Dissertations & Theses (Open Access)

Computer-controlled therapy machines allow for increasingly complex plans, as there are more variables that can be tuned to produce the ideal result. This makes it increasingly difficult to assure the intended calculated dose is being delivered correctly using current techniques that are 2D-based because the resultant dose distributions can differ markedly in various sections of the target. A measurement of composite dose from the entire plan should be included in patient-specific IMRT QA. A volumetric dosimeter such as PRESAGE® is able to provide a complete 3D measured dosimetry dataset with one treatment plan delivery. It was hypothesized that a …


An Implantable Mosfet Dosimeter Modified To Act As A Fiducial Marker, Joseph S. Dick Aug 2010

An Implantable Mosfet Dosimeter Modified To Act As A Fiducial Marker, Joseph S. Dick

Dissertations & Theses (Open Access)

In external beam radiation therapy, it is imperative that the prescribed dose is administered to the correct location and in the correct amount. Though several ex vivo methods of quality assurance are currently employed to achieve this goal, verifying that the correct dose is received within the patient in situ is impossible without the capability of measuring dose inside the patient. Recently, a method of measuring dose delivered within the patient has been developed, an implantable MOSFET dosimeter. This dosimeter is implanted within the patient and records the dose received. Since the dosimeter is implanted in the patient, it could …


Implementation Of An Anthropomorphic Phantom For The Evaluation Of Proton Therapy Treatment Procedures, Ryan L. Grant May 2010

Implementation Of An Anthropomorphic Phantom For The Evaluation Of Proton Therapy Treatment Procedures, Ryan L. Grant

Dissertations & Theses (Open Access)

With an increasing number of institutions offering proton therapy, the number of multi-institutional clinical trials involving proton therapy will also increase in the coming years. The Radiological Physics Center monitors sites involved in clinical trials through the use of site visits and remote auditing with thermoluminescent dosimeters (TLD) and mailable anthropomorphic phantoms. Currently, there are no heterogeneous phantoms that have been commissioned to evaluate proton therapy. It was hypothesized that an anthropomorphic pelvis phantom can be designed to audit treatment procedures (patient simulation, treatment planning and treatment delivery) at proton facilities to confirm agreement between the measured dose and calculated …


Commissioning An Anthropomorphic Spine And Lung Phantom For Remote Dose Verification Of Institutions Participating In Rtog 0631, Douglas F. Caruthers May 2010

Commissioning An Anthropomorphic Spine And Lung Phantom For Remote Dose Verification Of Institutions Participating In Rtog 0631, Douglas F. Caruthers

Dissertations & Theses (Open Access)

The RPC developed a new phantom to ensure comparable and consistent radiation administration in spinal radiosurgery clinical trials. This study assessed the phantom’s dosimetric and anatomic utility. The ‘spine phantom’ is a water filled thorax with anatomy encountered in spinal radiosurgery: target volume, vertebral column, spinal canal, esophagus, heart, and lungs. The dose to the target volume was measured with axial and sagittal planes of radiochromic film and thermoluminescent dosimeters (TLD). The dose distributions were measured with the radiochromic film calibrated to the absolute dose measured by the TLD. Four irradiations were administered: a four angle box plan, a seven …