Commissioning Of Micro-Cube Thermoluminescent Dosimeters For Small Field Dosimetry Quality Assurance In Radiotherapy, Brandon Luckett
Commissioning Of Micro-Cube Thermoluminescent Dosimeters For Small Field Dosimetry Quality Assurance In Radiotherapy, Brandon Luckett
Dissertations & Theses (Open Access)
Small field dosimetry presents complications and uncertainties that could be circumvented by using detectors which are smaller than the radiation field. This study evaluates the reproducibility and accuracy of TLD micro-cubes for use in stereotactic radiosurgery (SRS) remote auditing quality assurance (QA) for treatment centers participating in clinical trials. This study tested the hypothesis that TLD micro-cubes could be commissioned to evaluate small field dosimetry, and provide reproducibility within 3%, as well as assure agreement between measured dose and calculated doses to within 5%.
The aims of this thesis were to characterize and commission TLD micro-cubes as well as to …
High Frequency Percussive Ventilation For Tumor Motion Immobilization, Ina Marina Sala
High Frequency Percussive Ventilation For Tumor Motion Immobilization, Ina Marina Sala
Wayne State University Dissertations
This work investigates the use of High Frequency Percussive Ventilation as a technique for respiratory motion mitigation in radiotherapy. This technique was extensively investigated in several prospective and retrospective studies.
In an initial prospective study, we evaluated the feasibility of HFPV and chest-wall motion reduction, by recruiting 15 healthy volunteers to undergo HFPV with three commercially available interfaces. For direct tumor motion immobilization, a second prospective study was performed in which with ten lung cancer patients underwent HFPV while imaged with high frame rate fluoroscopy. Diaphragm motion and image artifacts were quantified in a prospective study of a healthy volunteer …
Utilizing Log Files For Treatment Planning And Delivery Qa In Radiotherapy, Carl W. Stanhope
Utilizing Log Files For Treatment Planning And Delivery Qa In Radiotherapy, Carl W. Stanhope
Wayne State University Theses
Purpose: Monte Carlo-based log file quality assurance (LF-MC QA) is investigated as an alternative method to phantom-based patient-specific quality assurance in radiotherapy (e.g. ArcCHECK QA (AC QA)).
Methods: First, the shortcomings of AC QA were investigated. The sensitivity dependence of ArcCHECK diodes on dose rate (in-field) and energy (primarily out-of-field) was quantified. LF-MC QA was then analyzed on the phantom geometry. Planned (‘Plan’) and LF-reconstructed CS and MC doses were compared with each other and AC measurement via statistical (mean ± StdDev(σ)) and gamma analyses to isolate dosimetric uncertainties and quantify the relative accuracies of AC QA and LF-MC QA. …