Open Access. Powered by Scholars. Published by Universities.®

Medicine and Health Sciences Commons

Open Access. Powered by Scholars. Published by Universities.®

Articles 1 - 10 of 10

Full-Text Articles in Medicine and Health Sciences

Development And Testing Of A Mapcheck3-Based Patient-Specific Qa Method For Prime Intensity Modulators Used With Bolus Electron Conformal Therapy, James T. Crist Jul 2023

Development And Testing Of A Mapcheck3-Based Patient-Specific Qa Method For Prime Intensity Modulators Used With Bolus Electron Conformal Therapy, James T. Crist

LSU Master's Theses

Purpose: Irregular surfaces in bolus electron conformal therapy (BECT) create target volume dose inhomogeneity, which off-axis intensity modulation (IM) can reduce. This research developed and tested a patient-specific clinical quality assurance (QA) procedure for fabricated Passive Radiotherapy Intensity Modulation for Electrons (PRIME) devices used for IM-BECT delivery.

Methods: To evaluate MapCheck3 (MC3) diode array accuracy, electron dose measurements beneath PRIME devices were compared with scanned measurements in a water phantom. Lateral dose profile comparisons were made for 35 combinations of five beam energies (7-20MeV) and seven PRIME devices - converging and diverging island blocks (pins) having 100-68.8% intensities. To evaluate …


Evaluation Of Residual Antibacterial Effects On Canine Skin Surface And Hairs Following Treatment With Five Commercial Mousse Products Against Staphylococcus Pseudintermedius In Vitro, Chi-Yen Wu Jul 2022

Evaluation Of Residual Antibacterial Effects On Canine Skin Surface And Hairs Following Treatment With Five Commercial Mousse Products Against Staphylococcus Pseudintermedius In Vitro, Chi-Yen Wu

LSU Master's Theses

Topical products are essential treatments for canine superficial bacterial folliculitis. However, studies for commercial products typically evaluate antibacterial effects on hairs rather than the skin surface. In addition, the interference from the hair length has not been evaluated in these studies. Therefore, the study objective for this study was to evaluate the residual antibacterial effects of five mousse products, (1) 2% chlorhexidine and 2% miconazole, (2) 0.05% phytosphingosine, (3) 2% salicylic acid and 10% ethyl lactate, (4) 3% chlorhexidine and 0.5% climbazole, and (5) 2% chlorhexidine and 1% ketoconazole on both the skin surface and hairs in both short- and …


In-Phantom Film Measurements Of Two Treatment Planning Systems For Single-Fraction Spine Sbrt, Michael J. Taylor May 2022

In-Phantom Film Measurements Of Two Treatment Planning Systems For Single-Fraction Spine Sbrt, Michael J. Taylor

LSU Master's Theses

Purpose: Treatment planning accuracy for spine stereotactic body radiation therapy (SBRT) varies depending on the dose calculation algorithm utilized in the treatment planning system (TPS). This project compared the end-to-end accuracy between spine SBRT plans calculated in a convolution-superposition based TPS (TPSCS) and Monte Carlo based TPS (TPSMC) with radiochromic film measurements. The hypothesis was that TPSMC would calculate the dose gradient in the critical region between the vertebral body and the spinal cord more accurately than TPSCS.

Methods: Single-fraction spine SBRT treatments following RTOG 0631 and local institutional guidelines were planned in …


Modeling Of Island Block Scatter For Intensity Modulated Bolus Electron Conformal Therapy, Joseph G. Scotto Jul 2021

Modeling Of Island Block Scatter For Intensity Modulated Bolus Electron Conformal Therapy, Joseph G. Scotto

LSU Master's Theses

Purpose: Bolus electron conformal therapy (BECT) benefits treatment of the post-mastectomy chest wall, head and neck, paraspinal muscles, and extremities. Patient dose heterogeneities caused by bolus can be reduced through intensity modulation (IM) across the incident electron beam. This requires passive radiotherapy intensity modulation for electrons (PRIME) devices, which utilize tungsten pins (island blocks) imbedded in machinable foam. IM-BECT treatment planning requires accurate dose calculation using the pencil beam redefinition dose algorithm (PBRA). Currently, the PBRA models island blocks as perfect collimators. This work explores models to account for electrons scattering into and out of the sides of island blocks. …


Structural Shielding Considerations For Vmat, Ana Lucia Dieguez Jul 2021

Structural Shielding Considerations For Vmat, Ana Lucia Dieguez

LSU Master's Theses

Introduction: As noted in National Council on Radiation Protection and Measurements (NCRP) Report 151, the medical physicist or other qualified expert has the responsibility to keep abreast of any new technology or treatment method that could potentially impact structural shielding design. Volumetric arc therapy (VMAT) became prevalent after the publication of Report 151 and thus was not explicitly addressed in Report 151. If the shielding-related characteristics of VMAT differ enough from the expectations of Report 151, especially in the circumstance of a vault utilized exclusively for VMAT, a shielding design based on Report 151 could potentially be inadequate. The …


Log File-Based Dose Reconstruction To Moving Targets During Lung Stereotactic Body Radiation Therapy, Andrew S. Mcguffey Jun 2020

Log File-Based Dose Reconstruction To Moving Targets During Lung Stereotactic Body Radiation Therapy, Andrew S. Mcguffey

LSU Master's Theses

Purpose: To perform film-based verification of 4D dose reconstruction to moving targets during lung stereotactic body radiation therapy (SBRT).

Introduction: Current patient-specific quality assurance measures to test deliverability of plans with dynamic intensity modulation involve delivering beams to static measurement device and comparing the planned dose to measurement. However, motion-induced dose errors are not detected with static measurement. Previous studies have investigated combining machine log data with respiratory tracking to determine moving-target dose. By combining machine log data with anatomic and density information at each breathing phase from 4D-CT, intrafraction anatomical deformation due to respiration may be accounted for. However, …


Analytical Setup Margin For Spinal Sbrt Based On Measured Errors, Audrey Copeland Jun 2020

Analytical Setup Margin For Spinal Sbrt Based On Measured Errors, Audrey Copeland

LSU Master's Theses

Purpose: No consensus currently exists in the radiotherapy community about the correct margin size to use for spinal SBRT. Margins have been proposed to account for various errors individually, but not with all errors combined to result in a single margin value. The purpose of this work was to determine a setup margin for spinal SBRT based on known and measurable errors during radiotherapy to achieve at least 90% coverage of the clinical target volume (CTV) with the prescription dose for at least 90% of patients and not exceed a 30 Gy point dose or 23 Gy to 10% of …


Scintillation Event Localization In Novel Hemi-Ellipsoid Detector For Spect, Hanif R. Soysal May 2019

Scintillation Event Localization In Novel Hemi-Ellipsoid Detector For Spect, Hanif R. Soysal

LSU Master's Theses

A high sensitivity Cardiac SPECT system using hemi-ellipsoid crystals with pinhole collimation was previously proposed by Dey et al. To investigate detector resolution for this design, the scintillation light spread on a monolithic hemi-ellipsoidal CsI crystal was simulated using Geant4 Monte Carlo. The expected distribution of scintillation light on the outer surface of the crystal from photoelectric absorption events was obtained from the simulations for selected points inside the crystal. Two sets of simulations were performed. For the first set, a look-up table (LUT) of 12 points was created and each point mapped to its expected light distribution—four points at …


Analytical Models Of Neutron Spectral Fluence, Kerma And Absorbed-Dose For Proton Therapy, Suman Shrestha Aug 2018

Analytical Models Of Neutron Spectral Fluence, Kerma And Absorbed-Dose For Proton Therapy, Suman Shrestha

LSU Master's Theses

Abstract

Purpose: The accurate prediction of stray neutron dose has become increasingly important as it increases the risk of second cancer development after proton therapy. Previously reported analytical models predicted the quantity dose equivalent, which includes physical and biological considerations but does not explicitly take into account material dependence and variation in the radiation quality. The purpose of this study was to investigate the feasibility of an analytical model of absorbed dose to water from stray neutrons in proton therapy.

Methods: To calculate neutron absorbed dose and kerma in water, the authors developed analytical models of neutron spectral fluence and …


Investigation Of Stereotactic Body Radiation Therapy Delivery Accuracy On An Elekta Linear Accelerator, Addie Barron Jul 2018

Investigation Of Stereotactic Body Radiation Therapy Delivery Accuracy On An Elekta Linear Accelerator, Addie Barron

LSU Master's Theses

Purpose: This work investigated the delivery accuracy of high-dose lung and spine stereotactic treatments delivered with the Elekta Infinity and Versa HD platforms. The accuracy of these platforms will be used for consideration in implementing a spine stereotactic radiosurgery (SSRS) program at Mary Bird Perkins Cancer Center.

Methods: A geometric phantom was used to perform Winston-Lutz type tests that assessed the relevant degrees of freedom (gantry, collimator, and couch) of the delivery system. A lung stereotactic body radiation therapy (SBRT) and spine SRS treatment plan were generated for use in end-to-end testing. Delivery accuracy was tested using a novel diode …