Open Access. Powered by Scholars. Published by Universities.®

Medicine and Health Sciences Commons

Open Access. Powered by Scholars. Published by Universities.®

Medical Specialties

Department of Radiation Oncology Faculty Papers

2017

Applicator

Articles 1 - 1 of 1

Full-Text Articles in Medicine and Health Sciences

Quality Assurance Guidelines For Superficial Hyperthermia Clinical Trials : Ii. Technical Requirements For Heating Devices., Hana Dobšíček Trefná, Johannes Crezee, Manfred Schmidt, Dietmar Marder, Ulf Lamprecht, Michael Ehmann, Jacek Nadobny, Josefin Hartmann, Nicolleta Lomax, Sultan Abdel-Rahman, Sergio Curto, Akke Bakker, Mark D. Hurwitz, Chris J. Diederich, Paul R. Stauffer, Gerard C. Van Rhoon May 2017

Quality Assurance Guidelines For Superficial Hyperthermia Clinical Trials : Ii. Technical Requirements For Heating Devices., Hana Dobšíček Trefná, Johannes Crezee, Manfred Schmidt, Dietmar Marder, Ulf Lamprecht, Michael Ehmann, Jacek Nadobny, Josefin Hartmann, Nicolleta Lomax, Sultan Abdel-Rahman, Sergio Curto, Akke Bakker, Mark D. Hurwitz, Chris J. Diederich, Paul R. Stauffer, Gerard C. Van Rhoon

Department of Radiation Oncology Faculty Papers

Quality assurance (QA) guidelines are essential to provide uniform execution of clinical trials with uniform quality hyperthermia treatments. This document outlines the requirements for appropriate QA of all current superficial heating equipment including electromagnetic (radiative and capacitive), ultrasound, and infrared heating techniques. Detailed instructions are provided how to characterize and document the performance of these hyperthermia applicators in order to apply reproducible hyperthermia treatments of uniform high quality. Earlier documents used specific absorption rate (SAR) to define and characterize applicator performance. In these QA guidelines, temperature rise is the leading parameter for characterization of applicator performance. The intention of this …