Prospective Comparison Of Point-Of-Care Device And Standard Analyzer For Monitoring Of International Normalized Ratio In Outpatient Oral Anticoagulant Clinic, Bushra Moiz, Ahmed Raheem, Anila Rashid, Muhammad Hasan, Lena Jafri Jan 2018

Prospective Comparison Of Point-Of-Care Device And Standard Analyzer For Monitoring Of International Normalized Ratio In Outpatient Oral Anticoagulant Clinic, Bushra Moiz, Ahmed Raheem, Anila Rashid, Muhammad Hasan, Lena Jafri

Department of Pathology and Laboratory Medicine

Point-of-care testing (POCT) coagulometers are increasingly being used in the hospital setting and patients’ self-testing. We determined the agreement of prothrombin time international normalized ratio (INR) results by POCT coagulometer and laboratory instrument through a comparative analysis and investigated whether the results of POCT coagulometer can reliably be used without being confirmed by standard laboratory analyzer. A total of 200 INR measurements by POCT coagulometer (CoaguChek XS Pro) and laboratory analyzer (Sysmex CS2000i) were compared using Passing-Bablok regression analysis and Bland-Altman plot. Agreement of the INR measurement was further analyzed in relation to dosing decision. The correlation of INR measurements …