Open Access. Powered by Scholars. Published by Universities.®
Articles 1 - 1 of 1
Full-Text Articles in Medicine and Health Sciences
Portal Imaging Using A Csi (Tl) Scintillator Coupled To A Cooled Ccd Camera, Amit Sawant
Portal Imaging Using A Csi (Tl) Scintillator Coupled To A Cooled Ccd Camera, Amit Sawant
Theses and Dissertations (ETD)
The purpose of this research was to design a high performance digital portal imaging
system, using a transparent x-ray scintillator coupled to a cooled CCD camera. Theoretical
analysis using Monte Carlo simulation was performed to calculate the QDE, SNR and DQE of
the system. A prototype electronic portal imaging device (EPID) was built, using a 12.7 mm
thick, 20.32 cm diameter, CsI (Tl) scintillator, coupled to an Astromed ® liquid nitrogen cooled
CCD TV camera. The system geometry of the prototype EPID was optimized to achieve high
spatial resolution. Experimental evaluation of the prototype EPID was performed, by
determining its …