An Evaluation Of The Host Response To An Interspinous Process Device Based On A Series Of Spine Explants: Device For Intervertebral Assisted Motion (Diam®), Jeffrey M. Toth, Justin D. Bric
An Evaluation Of The Host Response To An Interspinous Process Device Based On A Series Of Spine Explants: Device For Intervertebral Assisted Motion (Diam®), Jeffrey M. Toth, Justin D. Bric
Biomedical Engineering Faculty Research and Publications
Background:
The objective of this study was to evaluate the host response to an interspinous process device [Device for Intervertebral Assisted Motion (DIAM®)] based on a series of nine spine explants with a mean post-operative explant time of 35 months.
Methods:
Explanted periprosthetic tissues were processed for histology and stained with H&E, Wright-Giemsa stain, and Oil Red O. Brightfield and polarized light microscopy were used to evaluate the host response to the device and the resultant particulate debris. The host response was graded per ASTM F981-04. Quantitative histomorphometry was used to characterize particle size, shape, and area per ASTM F1877-05. …