Biomechanical Evaluation Of Locked And Non-Locked Constructs Under Axial And Torsion Loading, Vinit A. Patel
Biomechanical Evaluation Of Locked And Non-Locked Constructs Under Axial And Torsion Loading, Vinit A. Patel
Browse all Theses and Dissertations
Locking compression plates are proven to be safe for use in open reduction and internal fixation (ORIF). The ORIF is a procedure performed to treat fractures. It has various combination of holes, the system provides more options for clinicians to use either locking screw or non-locking screws. This thesis investigates and determines the best construct with special locking and non-locking screws under both axial and torsion loading. Twenty femur constructs were assembled with 2 cm osteotomy gap between femur shaft and condyle, bridged with 4.5 mm - 10 holes condyle plates. Femurs were divided in to 4 groups according to …