New In Vivo And In Vitro Techniques For Quantification Of Lateral Force Levels That Cause Implants To Fail : A Progress Report On Monkey Research, V. Leroy Leggitt
New In Vivo And In Vitro Techniques For Quantification Of Lateral Force Levels That Cause Implants To Fail : A Progress Report On Monkey Research, V. Leroy Leggitt
Loma Linda University Electronic Theses, Dissertations & Projects
Two new techniques for quantification of forces generated by an expansion screw apparatus placed between pairs of cylindrical endosseous titanium implants are described, along with preliminary data gained from their use in Macaca rhesus monkeys. Strain gages were bonded to the expansion apparatus and were calibrated either directly by micro load cell, or indirectly by Instron Machine to reflect the applied force.
Immediate loads of up to 6 kilograms have been applied to the implants with no increase in mobility as verified by a Periotest device. No previous studies on the orthodontic uses of implants have evaluated in vivo forces …