2014

Biomaterials

Implants

Articles 1 - 1 of 1

Full-Text Articles in Biomedical Engineering and Bioengineering

A Rubric For Electrochemical Testing Of Metallic Biomaterials, Frederick G. De La Fuente Aug 2014

A Rubric For Electrochemical Testing Of Metallic Biomaterials, Frederick G. De La Fuente

Master's Theses

Corrosion is a major factor for the failure of metallic medical implants. Testing a metal’s suseptability to corrosion prior to implantation is key to a successful implantation. Electrochemical processes were used in this study to evaluate the characteristics of corrosion of both AISI 316 stainless steel and titanium alloy Ti6Al4V, welded and non-welded. Linear, potentiodynamic, and cyclic polarization curves were produced by the PARC 2273 potentiostat showing the corrosion tendencies of the metals in four unique solutions 3.5% NaCl, 0.35% NaCl, phosphate buffered saline solution (PBS), and Butterfield phosphate buffered solution (BPS). The concentration of chloride ions in solutions affected …