Open Access. Powered by Scholars. Published by Universities.®

Education Commons

Open Access. Powered by Scholars. Published by Universities.®

Life Sciences

Series

Kinesiology, Health and Sport Studies

All-inside fixation devices

Articles 1 - 1 of 1

Full-Text Articles in Education

Shear Force At Failure And Stiffness Of All-Inside Meniscal Repair Devices, William K. Kesto, Amanda O. Esquivel, David C. Markel Dec 2013

Shear Force At Failure And Stiffness Of All-Inside Meniscal Repair Devices, William K. Kesto, Amanda O. Esquivel, David C. Markel

Kinesiology, Health and Sport Studies

The purpose of this study was to determine the failure load and stiffness of various meniscal repair devices. A total of 61 fresh-frozen porcine menisci (medial and lateral) were used for the study. A 30-mm vertical, full-thickness tear was created and repaired using one of three all-inside fixation devices and one inside–out repair in the vertical mattress pattern. We used the MaxBraid (Biomet, Warsaw, IN) inside–out suture as a control. The other devices tested were the Meniscal Cinch (Arthrex, Naples, FL), Ultra FasT-Fix (Smith & Nephew, Andover, MA), and the MaxFire MarXmen (Biomet, Warsaw, IN). In addition, two devices, MaxFire …