Open Access. Powered by Scholars. Published by Universities.®

Digital Commons Network

Open Access. Powered by Scholars. Published by Universities.®

Theses/Dissertations

Life Sciences

1997

Munster Technological University

Articles 1 - 1 of 1

Full-Text Articles in Entire DC Network

The Measurement Of Angiotensin-Converting Enzyme In Human Serum, Andrew P. Kenny Jan 1997

The Measurement Of Angiotensin-Converting Enzyme In Human Serum, Andrew P. Kenny

Theses

The measurement of angiotensin converting enzyme using the substrate FAPGG was investigated with a view to optimising its performance. It was found that a 100 mnol/l triethanolamine-HCl buffer gave a significantly higher absorbance change then the existing TRIS-HCl in the method. As this buffer has not been previously described in the measurement of ACE it was decided to introduce this buffer into a kinetic procedure and evaluate the modified method. The reference interval for the triethanolamine Method was determined by measuring the ACE activity on 200 serum from blood donors and a significant difference (P < 0.05) in the SACE activity of male and female donors was observed.

The modified method was found …