The Effects Of Field-Of-View And Patient Size On Ct Numbers From Cone-Beam Computed Tomography., Katrina Y T Seet, Arvand Barghi, Slav Yartsev, Jacob Van Dyk
The Effects Of Field-Of-View And Patient Size On Ct Numbers From Cone-Beam Computed Tomography., Katrina Y T Seet, Arvand Barghi, Slav Yartsev, Jacob Van Dyk
Oncology Publications
Cone-beam computed tomography (CBCT) is used for patient alignment before treatment and is ideal for use in adaptive radiotherapy to account for tumor shrinkage, organ deformation and weight loss. However, CBCT images are prone to artifacts such as streaking and cupping effects, reducing image quality and CT number accuracy. Our goal was to determine the optimum combination of cone-beam imaging options to increase the accuracy of image CT numbers. Several phantoms with and without inserts of known relative electron densities were imaged using the Varian on-board imaging system. It was found that CT numbers are most influenced by the selection …