Open Access. Powered by Scholars. Published by Universities.®
Articles 1 - 1 of 1
Full-Text Articles in Entire DC Network
A Continuous Dilution Calibration Technique For Flame Atomic-Absorption Spectrophotometry, Julian Tyson, J. M.H. Appleton
A Continuous Dilution Calibration Technique For Flame Atomic-Absorption Spectrophotometry, Julian Tyson, J. M.H. Appleton
Chemistry Department Faculty Publication Series
Calibration in atomic-absorption spectrometry is achieved by means of a concentration-gradient chamber using a single concentrated standard solution. Calibration is rapid and extends over the entire working concentration range of the analyte, irrespective of the shape of the conventionally obtained calibration curve. No curve-fitting approximations are involved. With well-designed apparatus, deviations are less than 1%.